Physical Safety Inspection Criteria

See EH&S physical safety criteria for laboratory inspections.

Inspection Criteria

Finding Script

Finding Description

Risk Level

Reference

Are machine guards and engineering safety controls installed and functioning?

Unguarded/unprotected equipment has the potential to cause severe workplace injuries. Manufacturer safety engineering controls (guards, safety switches, paneling, shielding, etc.), minimizes unnecessary risk and help prevent such injuries. All manufacturer installed safety engineering controls must not be removed and must be functioning as intended. Additional controls may be required depending on the nature of the equipment and the environment it is functioning in. Remove all equipment without the required safety controls or label "Out of Service" until those safety controls are repaired and/or installed. Depending on the nature of the equipment, repair options may be available through Facilities Management, licensed repair vendor or equipment manufacturer. To see if any machine will require additional safety engineering controls contact ehsrap@ucsd.edu.

High

Engineering Controls for Laboratory Safety

Do all industrial equipment and shop areas have posted safety instructions and have all operators been trained? 

Industrial equipment and shop areas must be posted with hazard warnings and safety rules. Machine shop equipment (such as mills, drill press, saws, lathes, forming machines, baling equipment, CNC, etc.) must adhere to the manufacturer's safety instructions. All users of this equipment must have documented training for the machine shop area and the safety procedures as specified in the equipment manuals they utilize. High-hazard equipment must have “safe-use” rules posted on or near the equipment or readily accessible to all users. The manufacturer's operating instructions, required PPE and emergency stop mechanisms must be accessible to address any additional site specific hazards or conditions. Contact ehsrap@ucsd.edu to conduct a job safety analysis to determine if additional hazards need to be addressed.

Medium

Research Safety: Physical Hazards

UCOP Shop Safety Manual

Is the lab area free of slip, trip or fall hazards?

The walking and working surfaces contained conditions that could lead to a slip, trip and/or a fall. Work surfaces shall be maintained in good repair, and protected against slips, trips, and falls. Remove the material or fix the condition that is contributing to the slip, trip or fall hazard. If the condition cannot be fixed immediately, warning signs (such as "Caution Wet Floor") must be posted until the hazard has been fixed. If a repair is required, options may be available through Facilities Management or by contacting a licensed repair vendor or manufacturer of the equipment.

Medium

Research Safety: Physical Hazards

Are sharps stored and transported safely with sharp edge unexposed?

Protect lab personnel from sharps injuries by storing and transporting sharps in protective material, such as cork, Styrofoam or a puncture-proof container when not in use (or left on the bench). Examples of sharps include needles, razor blades, scalpels, knives, syringes and sharp-pointed/dissecting forceps.

Medium

How to Manage Sharps

For labs that use an ultracentrifuge, is there a current rotor operation log?

High-speed rotors are prone to metal fatigue. Failure to discard rotors after a predetermined amount of use can result in serious damage from a mechanical failure (rotor disintegration, damaged equipment, damaged laboratory space, worker injury, etc.). The amount of damage from an ultracentrifuge, which can reach rotational speeds of up to 150,000 rpm, is catastrophic. To prevent this an operational log must be maintained for all ultracentrifuge rotors and maintained near or on the ultracentrifuge. These rotors must be changed and discarded before the manufacturer's recommended operational time has expired. If a log cannot be found for the ultracentrifuge the rotors must be changed and a new log started. Installed new rotors must be compatible with the machine they are going in and installed according to the manufacturer's specifications.

Medium

How to Use a Centrifuge Safely: Mechanical Failure