Biological Safety Inspection Criteria
See EH&S biological safety criteria for laboratory inspections.
|
Finding Script |
Finding Description |
Risk Level |
Reference |
|---|---|---|---|
|
Are appropriate and correctly labeled biohazardous waste bags used? |
Improper bags were used for biohazard waste and/or bag labels were incomplete, inaccurate or missing. Biohazardous waste must be stored in double red biohazard waste bags. Orange bags can never be used. Red biohazard waste bags must be marked by the manufacturer with the International Biohazard Symbol, and ASTM D1922, or ASTM D1709 markings, and must be labeled correctly. The correct labeling format for biohazardous waste is: UCSD, Street address, City, State, Zip, Building, Room, and 858-534-HELP. |
Medium |
Biohazardous Containers and Bags Address templates are available for download on the Laboratory Safety Labels page. Pre-labeled red biohazardous waste bags are available for purchase from UC San Diego Core Bio Services. |
|
If the lab has BSC(s) in use, are they certified annually, after repair, or relocation? |
Biosafety cabinet(s) (BSC(s)) in use is either past due or missing current certification, or is in need of repair. BSCs that are past due, in disrepair or uncertified cannot be used until the certification is completed. BSCs are primary engineering controls that provide personnel, environmental and research protection. Functioning BSCs are essential to workplace safety and are required to be certified annually, after repair or relocation. BSC(s) not certified or needing repair must be unplugged and have a sign posted "Do not use - Out of Service" indicating that the equipment is not safe for use. Resolution of this finding is complete after BSC service is complete and a copy of the new certification report or a photo of the new certification tag on the BSC is provided to EH&S. |
Medium |
See Biological Safety Cabinets: Overview for more information. |
|
Are the lab's BSC(s) free of open flame and/or a gas line connection (unless exempted in BUA/HCP)? |
Open flame burners and plumbed gas lines are not allowed in biosafety cabinets (BSC). Coordinate with Facilities Management to facilitate removal or shutting off of gas lines and/or remove open flame equipment. Note: in rare cases, exceptions may be granted with review by the Campus Fire Marshal, Department Chair and the Institutional Biosafety Committee (IBC) and approvals are documented in the BUA and associated HCPs. |
Medium |
Consult the Biological Safety Cabinets: Open Flames and Flammable Gas Policy for safer alternatives to using an open flame. |
|
Do all lab room signs meet requirements for communicating biological hazards present in the space? |
Lab room signs must include the biohazard symbol, biosafety level, name and phone number of responsible personnel, PPE requirements, general occupational health requirements and required procedures for entering and exiting the laboratory. |
Medium |
|
|
Is all equipment used for biohazardous materials (BSL-2 or higher) labeled with the universal biohazard symbol? |
Refrigerators, incubators and other equipment containing, or used for procedures involving, biohazardous agents must be labelled with the international biohazard symbol in laboratories performing work at Biosafety Level 2 (BSL-2 or higher). Contact ehsrap@ucsd.edu to obtain these labels. |
Medium |
|
|
Is the red biohazardous waste bag used in an appropriate secondary container that is rigid, leakproof, covered and labeled with biohazard symbols, and free of visible contamination? |
Red biohazardous waste bags must be used in a rigid, leakproof, covered container labeled with the word "Biohazard," the international biohazard symbol, and free of visible contamination. The biohazard label should be on the lid and all lateral sides of the secondary container. The lid must be in place except when the container is being accessed. |
Medium |
|
|
If the lab has a BSC(s), is research conducted properly and is BSC maintained free of clutter and disinfected appropriately? |
Laboratory personnel were observed using a biosafety cabinet (BSC) improperly, resulting in compromised protection of personnel, environment, and/or the research product. The work surface of a BSC was either cluttered, the front or back grill was partly covered, work procedures were observed that cause airflow disruption, and/or the unit was not disinfected before and after each use as required by good microbiological technique and NIH/CDC guidelines. Additionally, hazardous chemicals are not to be used in a BSC because the airflow recirculates within the room - a fume hood is the proper location for use of these chemicals. |
Medium |
See details for proper BSC use on Biological Safety Cabinets: Usage Guidelines. |
Questions? Contact the EH&S Research Assistance Program.